Thermometer calibrator - The 3000 series are small and lightweight heat sources, ideal for checking the accuracy of digital thermometers and temperature probes. The thermometer calibrators have a temperature range of +33 to +300°C with a resolution of 0.1°C.
These calibrators offer a high level of stability (±0.5°C) and a stabilisation time of five minutes. Plug it in, switch it on, set the verification temperature with the front panel buttons and insert your probe into the correct size well. Compare the temperature reading of your thermometer to the display and the difference is the error.

The 3001 dry-well will accept probe sizes Ø3.3, 4, 4.76 and 6.35mm.
The 3002 dry-well will accept probe sizes Ø3.3, 4.76, 6.35 and 9.6mm.
The 3003 dry-well will accept probe sizes Ø4.76 and 12.7mm.
• portable temperature heat source
• ideal for checking thermometer accuracy
|
specification |
3000 series dry wells |
|
range |
+33 to +300°C |
|
resolution |
0.1°C |
|
accuracy |
±0.5°C (+33 to +199.9°C) and ±1°C (+200 to +300°C) |
|
heating time |
ambient to +300°C after 10mins |
|
well depth |
100mm |
|
power |
230 volt AC (115 volt available) |
|
dimensions |
57 x 125 x 158mm |
|
weight |
900 grams |
|
a traceable certificate of calibration is included with each instrument |
|
Reference thermometer
The Reference thermometer is a high accuracy PT100
instrument that is supplied with a 5-point UKAS Certificate of Calibration. Each certificate indicates deviations from standards at various check points: -18°C, 0°C, +40°C, +70°C and +100°C. Special ranges may be certified by arrangement with the ETI UKAS calibration laboratory.
The Reference thermometer is ideal for comparison checking the calibration and accuracy of other thermometers and probes, when used in conjunction with a stable temperature heat or chill source
3004 dry-well heat source calibrator
The 3004 dry-well is a small and lightweight heat source with replaceable probe inserts. The unit has a temperature range of +33 to +300°C with a resolution of 0.1°C.
The 3004 has a high level of stability and a stabilisation time of five minutes. Set the temperature and insert your probe into the correct size well. Compare the temperature reading of your thermometer to the display temperature and the difference is the error.
accessories
N/A