The MicroCal 1 and 1 Plus thermocouple simulators help ensure that the frequent checking of thermometer accuracies is a routine operation. Both calibrators are designed to simulate a chosen temperature to test thermocouple type K, J, T, R, N, S and E thermometers without the need for specialised equipment or conversion tables. The MicroCal 1 Plus calibrator also measures and simulates temperature.
Both calibrators feature a custom 10mm LCD display with alpha-numeric display line to prompt the user when changing parameters. Selectable parameters include: °C/°F, auto power off - enable/disable, CJC - internal/external and display contrast adjustment.
Now available with re-chargeable battery option!
| 271-100 | MicroCal 1 simulator calibrator |
£
Ex VAT
|
||
| 271-101 | MicroCal 1 Plus simulator/thermometer |
£
Ex VAT
|
||
| 816-100 | thermocouple lead set (7 types) |
£
Ex VAT
|
||
| 830-205 | protective PVC boot |
£
Ex VAT
|
||
| 839-160 | 1.5 volt MN2400 (AAA) |
£
Ex VAT
|
||
|
839-400
|
ETI battery pack & charger (AAA) more info->
|
£
Ex VAT
|

Each MicroCal is supplied with a one metre PVC type K thermocouple lead with miniature connectors and a five point UKAS Certificate of Calibration which indicates deviations from standards at the various points.
Thermocouple type K - range -200 to +1372°C
Thermocouple type J - range -200 to +1200°C
Thermocouple type T - range -270 to + 400°C
Thermocouple type R - range 0 to +1768°C
Thermocouple type N - range -200 to +1300°C
Thermocouple type S - range 0 to +1768°C
Thermocouple type E - range -140 to +1000°C
• two models - simulator or simulator/thermometer
• 12 adjustable temperature points
• for frequent checking of thermometer accuracies
• tests thermocouple type K, J, T, R, N, S & E thermometers

|
specification |
MicroCal 1 & 1 Plus |
|
range |
see table in 'more info' tab - 12 adjustable presets |
|
accuracy |
±0.3°C |
|
battery |
2 x 1.5 volt MN2400 (AAA) |
|
battery life |
300 hours |
|
sensor type |
thermocouple type K, J, T, R, N, S or E (selectable) |
|
display |
custom LCD |
|
dimensions |
35 x 73 x 141mm |
|
weight |
220 grams |
|
a UKAS certificate of calibration is included with each instrument |
|
MicroCal 2 & 3 simulator thermometers
The MicroCal 2 and 3 thermocouple simulators help ensure that the frequent checking of thermometer accuracies is a routine operation. Both instruments are designed to simulate a chosen temperature, allowing standard K, J or T thermocouple thermometers to be tested or recalibrated simply and quickly, without the need for additional specialised equipment or conversion tables.Both models feature a custom 10mm LCD display with alpha-numeric display line to prompt the user when changing parameters. Selectable parameters include; °C/°F, auto power off - enable/disable, CJC - internal/external and display contrast adjustment.
The MicroCal 2 has 12 preset temperatures for type K thermocouple -20, -10, 0, +10, +30, +50, +100, +195, +250, +500, +800 and +1000°C, any of these temperatures can be modified and saved by the user. The factory default temperatures can be recalled at any time.
The MicroCal 3 has 23 fixed temperature points for type K thermocouple -100, -50, -20, -10, 0, +10, +20, +30, +40, +50, +60, +80, +100, +150, +195, +250, +300, +400, +500, +600, +800, +1000 and
+1200°C.
accessories
The protective PVC boot helps to protect instruments against accidental damage. An optional lead set is also available, that comprises of seven leads, one for each thermocouple type K, J, T, R, N, S and E. Each PVC lead is one metre long and incorporates two miniature thermocouple plugs. Both accessories are suitable for all types of the MicroCal simulators/thermometers.
| 830-205 protective PVC boot |
![]() |
£
Ex VAT
|
|||
| 816-100 thermocouple lead set (x7 ) |
£
Ex VAT
|
||||
| 839-160 1.5 volt MN2400 (AAA) |
![]() |
£
Ex VAT
|